描述
mytron
Type: T2240
Temperature cabinet, water-cooled
Technical data:
External dimensions: width 1435 mm, height 1955 mm, depth 2590 mm
Dimensions of the test chamber: width 1050 mm, height 1400 mm, depth 1600 mm
Volume of the test chamber: approx. 2240 l
Opening for cables:
2 x 100 mm right
2 x 100 mm left
Temperature range: – 40 °C to + 100 °C at 23 °C room temperature
Temperature stability over time: +/- 0.5 °K
Temperature stability over space: +/- 1.0 °K
Cooling: water cooling,
Inlet and outlet 1" internal thread
Cooling water shut off by solenoid valve when device is OFF
Fault display in the event of a refrigeration machine error
Cooling water consumption: Inlet temperature approx. 15 °C, approx. 15 l/min,
Drain temperature approx. 22 °C (with continuous cooling approx. 27 °C)
Mains connection: 3NPE AC 50 Hz 400 V, 32 A CEE plug
Gedpfx Afjvtl Hqe Asuh
Power consumption: approx. 6.5 kVA
Fuse protection: 3 x 20.0 AT, 3 x 10.0 AT
Lighting: 25 watts each on the left and right side of the test room
Exhaust air connection: DN 100 mm plug-in system
Safety device: electronic overtemperature protection, permanently set
Flue gas monitoring with horn
Viewing window: five-pane glazing 130 x 360 mm
Refrigerant: R 404 A approx. 10 kg
Weight: approx. 850 kg
Condition: used
Scope of delivery: (see picture)
(Changes and errors in the technical data, information is subject to change!)
We will be happy to answer any further questions you may have over the phone.
廣告是自動翻譯的,翻譯中可能會出現一些錯誤。
Type: T2240
Temperature cabinet, water-cooled
Technical data:
External dimensions: width 1435 mm, height 1955 mm, depth 2590 mm
Dimensions of the test chamber: width 1050 mm, height 1400 mm, depth 1600 mm
Volume of the test chamber: approx. 2240 l
Opening for cables:
2 x 100 mm right
2 x 100 mm left
Temperature range: – 40 °C to + 100 °C at 23 °C room temperature
Temperature stability over time: +/- 0.5 °K
Temperature stability over space: +/- 1.0 °K
Cooling: water cooling,
Inlet and outlet 1" internal thread
Cooling water shut off by solenoid valve when device is OFF
Fault display in the event of a refrigeration machine error
Cooling water consumption: Inlet temperature approx. 15 °C, approx. 15 l/min,
Drain temperature approx. 22 °C (with continuous cooling approx. 27 °C)
Mains connection: 3NPE AC 50 Hz 400 V, 32 A CEE plug
Gedpfx Afjvtl Hqe Asuh
Power consumption: approx. 6.5 kVA
Fuse protection: 3 x 20.0 AT, 3 x 10.0 AT
Lighting: 25 watts each on the left and right side of the test room
Exhaust air connection: DN 100 mm plug-in system
Safety device: electronic overtemperature protection, permanently set
Flue gas monitoring with horn
Viewing window: five-pane glazing 130 x 360 mm
Refrigerant: R 404 A approx. 10 kg
Weight: approx. 850 kg
Condition: used
Scope of delivery: (see picture)
(Changes and errors in the technical data, information is subject to change!)