温度测试室外部保护CTS
T-30/950 Ex/I
温度测试室外部保护
CTS
T-30/950 Ex/I
製造年份
2011
狀況
二手
位置
Borken 

顯示圖片
显示地图
机器数据
价格和位置
- 位置:
- Einsteinstraße 8a, 46325 Borken, DE
撥打
优惠详情
- 產品ID:
- A7538378
- 參考編號:
- 23117
- 最後更新:
- 於 27.06.2025
描述
CTS
Type: T-30/950 Ex/I
CTS Temperature Test Chamber, Explosion-Proof
Control – 32-bit controller
LCD control panel
Digital display for set and actual values
Numerical input of set values
Error messages in plain text
RS 232 interface
Free switching channels for product-side test specimen activation
Temperature Testing:
Temperature range: -30°C to +80°C
Temperature deviation: ±0.3 K over time
Temperature change rate:
Heating: 2 K/min
Cooling: 1.5 K/min down to -10°C
Thermal radiation: approx. 1000 W at +20°C
approx. 200 W at -25°C
Nitrogen inerting with residual oxygen monitoring
With nitrogen inerting, the test chamber is first purged with a large flow rate of N2 until the
o2 content drops below 3%. Then, it switches to a base flow amount (approx. 10% of the large flow rate).
The nitrogen flow for the base amount is measured and displayed using a flowmeter and monitored via a limit switch.
O2 monitoring continuously extracts gas from the test chamber via a flame arrestor and passes it over an O2 sensor which checks the residual oxygen content of the extracted air.
Afterwards, the gas is returned to the test chamber via a second flame arrestor.
Ijdpfxsihb Ams Af Aoih
The extraction pump is monitored by a volumetric flow meter. Additional details can be found in the Dräger user manual in the appendix.
The gas system is currently deactivated!
All technical data refer to an ambient temperature of +25°C, nominal voltage 400V, without test specimen, without irradiation, and without additional equipment.
Dimensions:
Test chamber volume: 950 liters
Test chamber dimensions: Height approx. 700 mm / Width approx. 900 mm / Depth approx. 1500 mm
External dimensions: Height approx. 2400 mm / Width approx. 1270 mm / Depth approx. 3125 mm
Operating data:
Electrical connection: 400V +6/-10 %, 3/N, 50 Hz
Nominal power: 21.6 kW
Rated current: 36.5 A
Protection class: IP 54
Condensate drain: non-pressurized, R 1/2" inner thread
Sound pressure level:
廣告是自動翻譯的,翻譯中可能會出現一些錯誤。
Type: T-30/950 Ex/I
CTS Temperature Test Chamber, Explosion-Proof
Control – 32-bit controller
LCD control panel
Digital display for set and actual values
Numerical input of set values
Error messages in plain text
RS 232 interface
Free switching channels for product-side test specimen activation
Temperature Testing:
Temperature range: -30°C to +80°C
Temperature deviation: ±0.3 K over time
Temperature change rate:
Heating: 2 K/min
Cooling: 1.5 K/min down to -10°C
Thermal radiation: approx. 1000 W at +20°C
approx. 200 W at -25°C
Nitrogen inerting with residual oxygen monitoring
With nitrogen inerting, the test chamber is first purged with a large flow rate of N2 until the
o2 content drops below 3%. Then, it switches to a base flow amount (approx. 10% of the large flow rate).
The nitrogen flow for the base amount is measured and displayed using a flowmeter and monitored via a limit switch.
O2 monitoring continuously extracts gas from the test chamber via a flame arrestor and passes it over an O2 sensor which checks the residual oxygen content of the extracted air.
Afterwards, the gas is returned to the test chamber via a second flame arrestor.
Ijdpfxsihb Ams Af Aoih
The extraction pump is monitored by a volumetric flow meter. Additional details can be found in the Dräger user manual in the appendix.
The gas system is currently deactivated!
All technical data refer to an ambient temperature of +25°C, nominal voltage 400V, without test specimen, without irradiation, and without additional equipment.
Dimensions:
Test chamber volume: 950 liters
Test chamber dimensions: Height approx. 700 mm / Width approx. 900 mm / Depth approx. 1500 mm
External dimensions: Height approx. 2400 mm / Width approx. 1270 mm / Depth approx. 3125 mm
Operating data:
Electrical connection: 400V +6/-10 %, 3/N, 50 Hz
Nominal power: 21.6 kW
Rated current: 36.5 A
Protection class: IP 54
Condensate drain: non-pressurized, R 1/2" inner thread
Sound pressure level:
廣告是自動翻譯的,翻譯中可能會出現一些錯誤。
您的广告已成功删除
发生错误